Monte Carlo modeling of the performance of a reflective tube absorption meter
- 20 October 1992
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 31 (30) , 6463-6468
- https://doi.org/10.1364/ao.31.006463
Abstract
A Monte Carlo model of the behavior of photons in a reflective tube absorption meter has been developed and used to investigate how well such a meter can measure the absorption coefficient in waters of different optical types. The apparent, i.e., measured, absorption coefficient (am) is always greater than the true absorption coefficient (a). The ratio am/a increases linearly with the ratio of scattering to absorption (b/a) at a rate that depends on the scattering phase function of the water. The excess attenuation is mainly due to the failure of forward-scattered photons to undergo reflection at the cylinder wall. To achieve the highest accuracy possible with the meter, one must maximize the reflectivity and detector acceptance angle. Performance is also improved by increasing the cylinder diameter.Keywords
This publication has 3 references indexed in Scilit:
- Volume scattering function, average cosines, and the underwater light fieldLimnology and Oceanography, 1991
- Monte Carlo study of the nature of the underwater light field in, and the relationships between optical properties of, turbid yellow watersMarine and Freshwater Research, 1981
- Laboratory Analysis of the Selective Absorption of Light by Sea Water*Journal of the Optical Society of America, 1939