Evaluation of heating patterns of microwave interstitial applicators using miniature electric field and fluoroptic temperature probes
- 1 January 1991
- journal article
- Published by Taylor & Francis in International Journal of Hyperthermia
- Vol. 7 (3) , 485-492
- https://doi.org/10.3109/02656739109005012
Abstract
The SAR patterns were determined for four commercially available microwave (915 MHz) interstitial applicators. Values of SAR were determined using a miniature (3 mm diameter) implantable isotropic electric field probe or a custom 0.25 mm diameter fluoroptic temperature probe. These are the smallest such probes that are currently available. Similar radial variation of SAR was found at the axial position of the gap in the outer conductor for each applicator. Electric field probe measurements are much faster and avoid some of the errors caused by the rapid spatial variation of SAR with interstitial applicators. The major limitation on the electric field probe is its size; it is larger than the applicators being tested.Keywords
This publication has 6 references indexed in Scilit:
- Thermal distribution studies of helical coil microwave antennas for interstitial hyperthermiaInternational Journal of Radiation Oncology*Biology*Physics, 1988
- The distribution of power and heat produced by interstitial microwave antenna arrays: II. The role of antenna spacing and insertion depthInternational Journal of Radiation Oncology*Biology*Physics, 1988
- RF Energy Deposition in a Heterogeneous Model of Man: Near-Field ExposuresIEEE Transactions on Biomedical Engineering, 1987
- Interstitial microwave antennas for thermal therapyInternational Journal of Hyperthermia, 1987
- Simulated biological materials for electromagnetic radiation absorption studiesBioelectromagnetics, 1987
- The Electric-Field Probe Near a Material Interface with Application to the Probing of Fields in Biological BodiesIEEE Transactions on Microwave Theory and Techniques, 1979