New Reflection Oximeter
- 1 April 1960
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 31 (4) , 401-403
- https://doi.org/10.1063/1.1716990
Abstract
It is experimentally shown that the ratio of light reflected by a nonhemolysed blood sample at two suitable wavelengths is a linear function of the oxygen saturation. A simple instrument to determine oxygen saturation in vitro, based on these findings, is described.Keywords
This publication has 3 references indexed in Scilit:
- The determination of the oxygenation of blood in vitro by using reflected lightAmerican Heart Journal, 1953
- Influence of Oxygen Saturation, Erythrocyte Concentration and Optical Depth Upon the Red and Near-Infrared Light Transmittance of Whole BloodAmerican Journal of Physiology-Legacy Content, 1951
- The Oximeter, an Instrument for Measuring Continuously the Oxygen Saturation of Arterial Blood in ManReview of Scientific Instruments, 1942