MACULAR THICKNESS MEASUREMENTS IN NORMAL EYES WITH TIME-DOMAIN AND FOURIER-DOMAIN OPTICAL COHERENCE TOMOGRAPHY
- 1 July 2009
- journal article
- research article
- Published by Wolters Kluwer Health in Retina
- Vol. 29 (7) , 980-987
- https://doi.org/10.1097/iae.0b013e3181a2c1a7
Abstract
Purpose: To compare macular thickness measurements using time-domain optical coherence tomography (OCT) and Fourier-domain OCT. Methods: Thirty-two eyes from 32 normal patients underwent complete ophthalmic evaluation. Macular scanning using the StratusOCT and the RTVue-100 OCT were performed for a total of three times each on the same visit. The average retinal thicknesses of the nine macular sectors as defined by the Early Treatment Diabetic Retinopathy Study, along with the foveal center point and macular volume, were recorded. The SD, the coefficient of variation, and the intraclass correlation coefficient were calculated for each parameter studied. Comparisons were made between the two OCTs in terms of retinal thicknesses measurements, their reproducibility, and macular regional differences. Correlations between retinal thickness and demographic variables (age and gender) were also investigated. Due to known differences in segmentation algorithms of the two OCTs, software calipers were used to measure the distance from the internal limiting membrane to the photoreceptor inner segment–outer segment junction at the foveal center point on all RTVue scans to allow a more fair comparison. Results: The RTVue yielded greater retinal thickness measurements in nearly all macular subfields compared with the StratusOCT. Even after accounting for differences in segmentation algorithms, significant disparities were still evident with the RTVue measurements less than those of the StratusOCT at the foveal center. On both machines, the macula was thinnest at the fovea and thickest within the 3 mm ring. There were some consistent regional variations in macular thickness evident on both OCTs. Compared with the StratusOCT, the RTVue generally had lower coefficients of variation and higher intraclass coefficients, suggesting better reproducibility. Age and gender also appeared to be important determinants in some macular thickness parameters. Conclusion: Compared with StratusOCT, the RTVue fourier-domain OCT yields greater retinal thickness measurements with greater reproducibility, presumably due to different segmentation algorithms, increased sampling density, and greater resolution. However, regional differences across the macula can be consistently observed with both devices.Keywords
This publication has 36 references indexed in Scilit:
- Racial Differences in Macular Thickness in Healthy EyesOptometry and Vision Science, 2007
- Error Correction and Quantitative Subanalysis of Optical Coherence Tomography Data Using Computer-Assisted GradingInvestigative Opthalmology & Visual Science, 2007
- High-Definition and 3-dimensional Imaging of Macular Pathologies with High-speed Ultrahigh-Resolution Optical Coherence TomographyOphthalmology, 2006
- Histologic Correlation of In Vivo Optical Coherence Tomography Images of the Human RetinaAmerican Journal of Ophthalmology, 2006
- Quantitative In Vivo Retinal Thickness Measurement in Chinese Healthy Subjects with Retinal Thickness AnalyzerInvestigative Opthalmology & Visual Science, 2006
- Heritability of Macular Thickness Determined by Optical Coherence TomographyInvestigative Opthalmology & Visual Science, 2006
- Comparison of optical coherence tomography models OCT1 and Stratus OCT for macular retinal thickness measurementBritish Journal of Ophthalmology, 2005
- Comparison of Ultrahigh- and Standard-Resolution Optical Coherence Tomography for Imaging Macular PathologyOphthalmology, 2005
- Estudio de parámetros de normalidad en relación a la edad con el Tomógrafo Óptico de CoherenciaArchivos de la Sociedad Española de Oftalmología, 2005
- Macular and Retinal Nerve Fiber Layer Thickness Measurement Reproducibility Using Optical Coherence Tomography (OCT-3)Journal of Glaucoma, 2004