Centration of excimer laser photorefractive keratectomy relative to the pupil
- 1 January 1993
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 19 (Supplement) , 144-148
- https://doi.org/10.1016/s0886-3350(13)80398-7
Abstract
The centration of excimer laser photorefractive keratectomy (PRK) is critical to the procedure's success. We evaluated PRK centration in 49 patients using the EyeSys topography system. Ablation zone centration was measured from the corneal vertex and from the pupillary center using the pupil-finding software. Centration was measured more accurately from the pupillary center (0.40 mm) than from the corneal vertex (0.44 mm). Right eyes were decentered less than left eyes. There was an unpredictable correlation between amount of decentration and postoperative visual acuities. The ability to measure centration of keratorefractive procedures precisely from the pupil is an important advance in topography technology.Keywords
This publication has 10 references indexed in Scilit:
- Corneal topography following excimer photorefractive keratectomy for myopiaJournal of Cataract & Refractive Surgery, 1993
- Centration of excimer laser photorefractive keratectomy relative to the pupilJournal of Cataract & Refractive Surgery, 1993
- Pupil diameter and the principal rayJournal of Cataract & Refractive Surgery, 1992
- Changes in Corneal Topography after Excimer Laser Photorefractive Keratectomy for MyopiaOphthalmology, 1991
- Mean Visual AcuityAmerican Journal of Ophthalmology, 1991
- Corneal Topography and Optical Zone Location in Photorefractive KeratectomyJournal of Refractive Surgery, 1990
- The effect of mydriasis on the pupillary centration of the human eyeOphthalmic and Physiological Optics, 1988
- Corneal ModelingCornea, 1988
- Centering Corneal Surgical ProceduresAmerican Journal of Ophthalmology, 1987
- Terminology in Ocular Motility and Allied SubjectsAmerican Journal of Ophthalmology, 1943