Reproducibility of Normal Corneal Power Measurements With a Keratometer, Photokeratoscope, and Video Imaging System
- 1 April 1990
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 108 (4) , 539-544
- https://doi.org/10.1001/archopht.1990.01070060087055
Abstract
• To determine the reproducibility of currently available instruments for measuring corneal power, two investigators measured comparable locations on 18 normal human corneas using a keratometer (Bausch & Lomb, Rochester, NY), the Corneascope (Kera Corp, Santa Clara, Calif), and the Corneal Modeling System (Computed Anatomy Inc, New York, NY). (For the two keratoscopes, average powers around rings at comparable locations were used.) Comparisons made between instruments at the significance level of .05 indicated that the keratometer was more reproducible than the Corneal Modeling System and the Corneascope, and the Corneal Modeling System was more reproducible than the Corneascope. Of the 31 rings projected on the surface of each cornea by the Corneal Modeling System, rings 2 through 13 were read reasonably reproducibly in that 76% of the measurements on these rings differed by no more than 0.5 diopter. The Corneal Modeling System (software version 1.16) is 83% as reproducible as a keratometer reading at approximately the edge of the 3-mm central zone on normal human corneas and provides information about corneal topography in a more reproducible and visually useful manner than the other two instruments.This publication has 8 references indexed in Scilit:
- Corneal Topography of Early KeratoconusAmerican Journal of Ophthalmology, 1989
- Accuracy and Precision of Keratometry, Photokeratoscopy, and Corneal Modeling on Calibrated Steel BallsArchives of Ophthalmology (1950), 1989
- Irregular Astigmatism After Radial and Astigmatic KeratotomyAmerican Journal of Ophthalmology, 1989
- Topographic Analysis and Visual Acuity After Radial KeratotomyAmerican Journal of Ophthalmology, 1988
- Corneal ModelingCornea, 1988
- Corneoscopy in Keratorefractive SurgeryCornea, 1983
- Method for Calculation of Corneal Profile and Power DistributionArchives of Ophthalmology (1950), 1981
- Corneal topographyDocumenta Ophthalmologica, 1977