Reproducibility problems with the Abbott laboratories LCx assay for Chlamydia trachomatis and Neisseria gonorrhoeae.
- 1 June 2000
- journal article
- clinical trial
- Vol. 38 (6) , 2416-8
Abstract
This study demonstrates that significant reproducibility problems can occur during routine use of the Abbott Laboratories LCx assay for Chlamydia trachomatis and Neisseria gonorrhoeae. These problems can go undetected by the quality control procedures outlined in the manufacturer's package insert. We outline here procedures for detecting and preventing contamination and reproducibility problems.This publication has 9 references indexed in Scilit:
- Urine Specimens from Pregnant and Nonpregnant Women Inhibitory to Amplification of Chlamydia trachomatis Nucleic Acid by PCR, Ligase Chain Reaction, and Transcription-Mediated Amplification: Identification of Urinary Substances Associated with Inhibition and Removal of Inhibitory ActivityJournal of Clinical Microbiology, 1998