Thermoacoustic Molecular Imaging of Small Animals
Open Access
- 1 April 2003
- journal article
- Published by SAGE Publications in Molecular Imaging
- Vol. 2 (2) , 113-123
- https://doi.org/10.1162/153535003322331993
Abstract
We have designed, constructed, and tested a thermoacoustic computed tomography (TCT) scanner for imaging optical absorption in small animals in three dimensions. The device utilizes pulsed laser irradiation (680-1064 nm) and a unique, 128-element transducer array. We quantified the isotropic spatial resolution of this scanner to be 0.35 mm. We describe a dual-wavelength subtraction technique for isolating optical dyes with TCT. Phantom experiments demonstrate that we can detect 5 fmol of a near-infrared dye (indocyanine green, ICG) in a 1-microL volume using dual-wavelength subtraction. Initial TCT imaging in phantoms and in two sacrificed mice suggests that three-dimensional, optical absorption patterns in small animals can be detected with an order of magnitude better spatial resolution and an order of magnitude better low-contrast detectability in small animals when compared to fluorescence imaging or diffusion optical tomography.Keywords