Microseparation Chips for Performing Multienzymatic Dehydrogenase/Oxidase Assays: Simultaneous Electrochemical Measurement of Ethanol and Glucose
- 8 February 2001
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 73 (6) , 1296-1300
- https://doi.org/10.1021/ac001205t
Abstract
This report describes a new “lab-on-a-chip” protocol integrating on-line precolumn biocatalytic reactions of multiple (oxidase and dehydrogenase) enzymes and substrates with effective capillary electrophoresis microseparations and amperometric detection. The operation of the new oxidase/dehydrogenase reaction/separation microchip is illustrated for the simultaneous measurement of glucose and ethanol in connection to the corresponding glucose oxidase and alcohol dehydrogenase reactions, respectively. The enzymatic reactions generate hydrogen peroxide and NADH species that are separated (on the basis of their different charges) and detected amperometrically at the end-column thick-film detector. A driving voltage of 2000 V results in peroxide and NADH migration times of 74 and 230 s, respectively. Operating the gold-coated carbon detector at +1.0 V allows simultaneous anodic detection of both reaction products. Factors influencing the reaction, separation, and detection processes are examined and optimized. The applicability of the new multienzyme assay to wine samples is illustrated.Keywords
This publication has 12 references indexed in Scilit:
- Capillary electrophoresis microchips with thick-film amperometric detectors: separation and detection of phenolic compoundsAnalytica Chimica Acta, 2000
- Lab-on-a-Chip: A Revolution in Biological and Medical Sciences.Analytical Chemistry, 2000
- Miniaturized total analysis systems for biological analysisAnalytical and Bioanalytical Chemistry, 2000
- Micromachined Electrophoresis Chips with Thick-Film Electrochemical DetectorsAnalytical Chemistry, 1999
- Microfluidic Assays of Acetylcholinesterase InhibitorsAnalytical Chemistry, 1999
- Capillary Electrophoresis Chips with Integrated Electrochemical DetectionAnalytical Chemistry, 1998
- Amperometric biosensors based on NAD(P)‐dependent dehydrogenase enzymesElectroanalysis, 1997
- Integrated Capillary Electrophoresis Devices with an Efficient Postcolumn Reactor in Planar Quartz and Glass ChipsAnalytical Chemistry, 1996
- Precolumn Reactions with Electrophoretic Analysis Integrated on a MicrochipAnalytical Chemistry, 1994
- Native protein separations and enzyme microassays by capillary zone and gel electrophoresisAnalytical Chemistry, 1993