Acute Myocardial Infarction Secondary to Thromboembolism from a Fractured Prosthetic Aortic Valve
- 1 October 1968
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 38 (4) , 672-677
- https://doi.org/10.1161/01.cir.38.4.672
Abstract
A case of coronary artery embolism from thrombus in a fracture in an aortic Starr-Edwards prosthesis was documented at necropsy. This 40-year-old white man had replacement of the aortic and mitral valves by ball-valve prostheses 5 years prior to his death. Following surgery, he was able to return to full-time work. He did not return for follow-up for 2 years, and when next seen he had an acute myocardial infarction and left ventricular failure. An embolus to the left anterior descending coronary artery was demonstrated at the postmortem examination. The mitral ball was intact, but the aortic ball had two large fractures which contained thrombi.Keywords
This publication has 15 references indexed in Scilit:
- Successful Replacement of Defective Ball of a Prosthetic Aortic ValveJAMA, 1968
- Teflon embolism of coronary arteriesThe Journal of Thoracic and Cardiovascular Surgery, 1967
- Death Due to Swelling of Ball Component of Aortic Ball-Valve ProsthesisNew England Journal of Medicine, 1967
- Phonocardiogram of the ball-and-cage aortic valve prosthesis.Heart, 1967
- Factors Affecting Thromboembolism Associated with Prosthetic Heart ValvesCirculation, 1967
- Late Postoperative Pathological Findings After Cardiac Valve ReplacementCirculation, 1967
- Functional results with aortic ball valve prostheses (Starr-Edwards) followed for two to three years.Thorax, 1967
- Sutureless Mitral Valve ReplacementThe Annals of Thoracic Surgery, 1966
- Expulsion of Ball from Aortic Valve ProsthesisThe Annals of Thoracic Surgery, 1966
- Special Problems Following Valve Replacement: Embolus, Leak, Infection, Red Cell DamageThe Annals of Thoracic Surgery, 1965