Background Illumination and Automated Perimetry
- 1 March 1986
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 104 (3) , 395-397
- https://doi.org/10.1001/archopht.1986.01050150095035
Abstract
• Visual field function in the right and left eyes of 31 normal volunteers was evaluated with an automated projection perimeter (OCTOPUS). Serial visual field evaluations were repeated in these same eyes with neutral filters of increasing optical density. We compared the results of threshold determinations with the different neutral filters in place before the examined eye. Significant reduction in threshold sensitivity at several test spots throughout the central 30° visual field was seen with neutral density filters of 0.5 log units or greater. The low level of background illumination of the OCTOPUS perimeter (4.0 apostilbs) may allow relatively minor reduction in light transmission by the ocular media to produce significant changes in the recorded level of threshold sensitivity during visual field evaluation.This publication has 5 references indexed in Scilit:
- Quantitative Office PerimetryOphthalmology, 1985
- Effectiveness of Automated Perimetry in Following Glaucomatous Visual Field ProgressionOphthalmology, 1982
- Automated Suprathreshold Static PerimetryAmerican Journal of Ophthalmology, 1980
- Perimetry in Cataract PatientsArchives of Ophthalmology (1950), 1978
- Visual field defect in glaucoma.BMJ, 1971