Shielding Requirements for Mammography
- 31 August 1987
- journal article
- research article
- Published by Wolters Kluwer Health in Health Physics
- Vol. 53 (3) , 267-279
- https://doi.org/10.1097/00004032-198709000-00006
Abstract
Shielding requirements for mammography installations have been investigated. To apply the methodologies of NCRP Report No. 49, the scatter-to-incident ratio of a typical mammography beam was measured, and the broad beam transmission was calculated for several representative beam spectra. These calculations were found to compare favorably with published low kVp tungsten-targeted X-ray transmission through a variety of shielding materials. Radiation shielding tables were developed from the calculated transmissions through Pb, concrete, gypsum, steel, plate glass, and water, using a technique which eliminates the "add one HVL" rule. It is concluded that Mo-targeted X-ray beams operated at 35 kVp require half the shielding of W-targeted beams operated at 50 kVp, and that adequate, cost-effective shielding calculations will consider alternatives to Pb.Keywords
This publication has 2 references indexed in Scilit:
- Radiation Shielding of Multiple X-ray Sources in Diagnostic RadiologyHealth Physics, 1986
- Use of Gypsum Wallboard For Diagnostic X-ray Protective BarriersHealth Physics, 1979