New method for improving accuracy of 24-hour continuous intragastric pH-metry
- 1 July 1994
- journal article
- research article
- Published by Springer Nature in Digestive Diseases and Sciences
- Vol. 39 (7) , 1416-1424
- https://doi.org/10.1007/bf02088043
Abstract
Continuous 24-hr intragastric pH-metry was prospectively performed in 801 subjects with different clinical conditions using two pH electrodes placed closely adjacent. The aim was to assess thein situ repeatability of the test and to verify whether the removal of artifacts, interference, and noise usually superimposed onto the fundamental signal recorded by the measuring apparatus improves the clinical usefulness of experimental data. The following debugging/filtering procedure was used: first, pH recordings of each channel were amended separately from artifacts, then they underwent 7 min windowed median interference debugging, and finally Wiener noise filtering was applied. Afterwards, the 24-hr mean pH profile was obtained in each subject by averaging the pH tracings of the two channels every minute (1440 data points/24 hr). The efficiency of this procedure was assessed at each step by evaluating the difference among groups using the O'Brien test, a distribution-free nonparametric method well-suited for evaluating differences among groups allocated onto a two-way layout. The differences among groups calculated from raw pH data of the single channels can be very misleading, in that it is possible to find that they are significant on one channel and not significant on the other channel. Conversely, the significance of the differences among groups increases progressively at each step of the above debugging/filtering procedure applied to raw pH profiles of each channel. Seven minutes was shown to be the most suitable time lag for windowed median removal of interference. The whole procedure increases by 10 up to 108 times the probability of detecting differences among groups, particularly when one rules out pH tracings with a discrepancy time between the two channels which is more than 1 hr over the entire 24-hr period. Our results suggest that removal of artifacts, interference, and noise superimposed onto the endoluminal pH tracings can greatly improve the usefulness of 24-hr intragastric pH-metry in every clinical situation. Moreover, test reliability largely depends on thein situ reproduceability of pH measurements.Keywords
This publication has 23 references indexed in Scilit:
- Variations in results of simultaneous ambulatory esophageal pH monitoringDigestive Diseases and Sciences, 1992
- The ion sensitive field effect transistor (ISFET) pH electrode: a new sensor for long term ambulatory pH monitoring.Gut, 1991
- Comparison of Gastric Aspiration and Continuous pH Monitoring with Antimony Electrode: Methodological RemarksDigestive Diseases, 1990
- Gastric Aspiration versus Antimony and Glass pH Electrodes: A Simultaneous Comparative in Vivo StudyScandinavian Journal of Gastroenterology, 1989
- Statistical Considerations for Performing Multiple Tests in a Single Experiment. 5. Comparing Two Therapies With Respect to Several EndpointsMayo Clinic Proceedings, 1988
- Pattern of 24 hour intragastric acidity in active duodenal ulcer disease and in healthy controls.Gut, 1988
- Does smoking interfere with the effect of histamine H2-receptor antagonists on intragastric acidity in man?Gut, 1987
- Continuous 24 hour ambulatory monitoring of intragastric pH in manClinical Physics and Physiological Measurement, 1987
- Long-term recording and computer analysis of simultaneous gastric and duodenal pH under normal conditions in manClinical Physics and Physiological Measurement, 1980
- The histological diagnosis of chronic gastritis in fibreoptic gastroscope biopsy specimensJournal of Clinical Pathology, 1972