Laser-Induced Electron Capture Mass Spectrometry

Abstract
Two techniques are reported for detection of electrophore-derivatized compounds by laser-induced electron capture time-of-flight mass spectrometry (LI-EC-TOF-MS). In both cases, a nitrogen laser is used to induce the electron capture. The analyte is deposited in a matrix consisting of a compound with a low ionization potential such as benzo[ghi]perylene in the first technique, where the electron for electron capture apparently comes from this matrix. In the second technique, the analyte is deposited on a silver surface in the absence of matrix. It seems that “monoenergetic” ions instantly desorb from the target surface in the latter case, since the peak width in the continuous extraction mode essentially matches the pulse width of the laser (4 ns). Ten picomoles of 3-O-(pentafluorobenzyl)-α-estradiol were detected at a S/N ≥ 50, where the spot size of the laser was ∼0.25% of the sample spot. It is attractive that simple conditions can enable sensitive detection of electrophores on routine TOF-MS equipment. The technique can be anticipated to broaden the range of analytes in both polarity and size that can be detected by EC-MS relative to the range for GC/EC-MS.