Reducing the effects of scattered photons in X-ray projection imaging
- 1 March 1989
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 8 (1) , 56-63
- https://doi.org/10.1109/42.20362
Abstract
An iterative algorithm is introduced which can compute and subtract the scatter bias. It is assumed that the scatter-free image is spatially bounded and anything outside the boundary is the result of recorded scattered photons. This information and the fact that the scatter distribution does not have appreciable high-frequency content are utilized.Keywords
This publication has 5 references indexed in Scilit:
- Phase-only signal reconstructionPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2005
- An algorithm to find two-dimensional signals with specified zero crossingsIEEE Transactions on Acoustics, Speech, and Signal Processing, 1987
- A technique of scatter and glare correction for videodensitometric studies in digital subtraction videoangiography.Radiology, 1982
- Deconvolution Technique For The Improvement Of Contrast Of Image IntensifiersPublished by SPIE-Intl Soc Optical Eng ,1981
- Scatter Considerations in Fan Beam Computerized Tomographic SystemsIEEE Transactions on Nuclear Science, 1976