Resistance and additional inspiratory work imposed by the laryngeal mask airway
- 1 April 1992
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 47 (4) , 343-347
- https://doi.org/10.1111/j.1365-2044.1992.tb02179.x
Abstract
Laryngeal mask airways and tracheal tubes were studied to determine both their resistance to constant gas flows and additional inspiratory work during simulated inspiration. Laryngeal mask airways imposed less resistance and required lower additional inspiratory work compared with the corresponding sized tracheal tubes. If inspiratory loading during anaesthesia is an important consideration, then the laryngeal mask airway may be preferable to a tracheal tube.Keywords
This publication has 18 references indexed in Scilit:
- Cardiovascular response to insertion of Brain's laryngeal maskAnaesthesia, 1990
- Additional work of breathing imposed by endotracheal tubes, breathing circuits, and intensive care ventilatorsCritical Care Medicine, 1989
- The laryngeal mask airwayAnaesthesia, 1989
- A rococo reportPublished by American Medical Association (AMA) ,1977
- Airflow resistances of endotracheal tubesJAMA, 1977
- Weaning from mechanical ventilationJAMA, 1976
- The ability of normal subjects to tolerate added inspiratory loadsRespiration Physiology, 1970
- A NEW METHOD FOR ANALYSIS OF RESPIRATORY WORK BY MEASUREMENTS OF THE ACTUAL POWER AS A FUNCTION OF GAS FLOW, PRESSURE AND TIME: A Preliminary ReportActa Anaesthesiologica Scandinavica, 1962
- THE RESPIRATORY EFFECTS OF RESISTANCE TO BREATHING IN ANESTHETIZED MANAnesthesiology, 1961