An analysis and optimization of elliptical RF probes used in magnetic resonance imaging
- 1 September 1996
- journal article
- Published by IOP Publishing in Measurement Science and Technology
- Vol. 7 (9) , 1281-1290
- https://doi.org/10.1088/0957-0233/7/9/013
Abstract
In magnetic resonance imaging of the entire body, it is often desirable to use an elliptical RF probe, rather than a circular one. As an ellipse more closely conforms to the anatomical cross section of the human thorax and head, better filling factors and therefore improved signal-to-noise ratios may be achieved by the use of elliptical RF coils. The probe is usually of bird-cage type, but the rungs are of finite width due to the high-frequency signals involved. This paper presents a method for computing the magnetic fields produced inside elliptical probes, and the current distributions on the rungs. A slotted shield is assumed to surround the probe, and its influence on field homogeneity is studied. In particular, the currents in a 16-runged unshielded elliptical coil of practical interest were determined optimally in one case, using simulated annealing to optimize the homogeneity of the magnetic field within the probe. The effects of a segmented shield of both elliptical and circular cross section on this coil are discussed, and the results are confirmed by experiment.Keywords
This publication has 13 references indexed in Scilit:
- Currents and fields in shielded RF resonators for NMR/MRIMeasurement Science and Technology, 1996
- Optimized Small-Bore, High-Pass Resonator DesignsJournal of Magnetic Resonance, Series B, 1995
- A Novel Approach to the Calculation of RF Magnetic and Electric Fields for NMR Coils of Arbitrary GeometryJournal of Magnetic Resonance, Series B, 1995
- Determining current distributions for RF resonators in magnetic resonance imagingMeasurement Science and Technology, 1995
- Gradient-Coil Design by Simulated AnnealingJournal of Magnetic Resonance, Series A, 1993
- A multiple-frequency coil with a highly uniform B1 fieldJournal of Magnetic Resonance (1969), 1989
- Currents and fields of thin conductors in rf saddle coilsMagnetic Resonance in Medicine, 1986
- An efficient, highly homogeneous radiofrequency coil for whole-body NMR imaging at 1.5 TJournal of Magnetic Resonance (1969), 1985
- Split-ring resonator for use in magnetic resonance from 200–2000 MHzReview of Scientific Instruments, 1981
- An efficient decoupler coil design which reduces heating in conductive samples in superconducting spectrometersJournal of Magnetic Resonance (1969), 1979