Fully automated interferometric PMD measurements for active EDFA, fibre optic components and optical fibres
- 2 September 1993
- journal article
- Published by Institution of Engineering and Technology (IET) in Electronics Letters
- Vol. 29 (18) , 1649-1651
- https://doi.org/10.1049/el:19931098
Abstract
Using an interferometric detection technique, highly accurate and fully automated polarisation mode dispersion (PMD) measuring equipment with 1.7fs (00017ps) accuracy has been developed. PMD measurements of the active EDFA, fibre optic components, and 100 km long optical fibres are presented. A comparison of the interferometric method with the Jones matrix method and the fixed analyser method is also described.Keywords
This publication has 0 references indexed in Scilit: