Changes in long‐bone structural properties during the first 8 weeks of plate implantation
- 1 January 1984
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 2 (1) , 80-89
- https://doi.org/10.1002/jor.1100020113
Abstract
The midplate structural rigidities of metal- and plastic-plated intact canine femora were experimentally determined after initial plate application and 8 weeks after in vivo implantation. Composite beam theory significantly overestimated the bending rigidities of the metal-plated bones. The rigidities of the plastic-plated bones were nearly identical to that of the isolated bone, as composite beam theory predicted. Plating with either plate increased the intracortical porosity and caused the deposition of periosteal new bone, which was greater with plastic than with metal plates. The increased rigidities provided by the attachment of the metal plates and, hence, the degree of bone strain shielding were variable. Platings for 8 weeks which provided little strain shielding with either metal or plastic plates caused an increase in bone flexural rigidity (measured after plate removal) with respect to the contralateral control. Platings that provided increasing amounts of strain shielding caused a decreasing midplate bone rigidity (measured after plate removal) and increasing bone deposition at the outer screws. These findings suggest that the surgical implantation of any plate (metal or plastic) will provide a net stimulus to bone formation and consequently increased bone structural rigidity, even though intracortical porosity is increased. If the plate significantly reduces the normal loads borne by the bone, however, there is a net stimulus to remove bone, resulting in a loss of midplate structural rigidity within 8 weeks of implantation.Keywords
This publication has 30 references indexed in Scilit:
- Stress analysis of a simplified compression plate fixation system for fractured bonesComputers & Structures, 1983
- A new approach to the design of internal fixation platesJournal of Biomedical Materials Research, 1983
- Experimental determination of whole long bone sectional propertiesJournal of Biomechanics, 1982
- In vitro Evaluation of the Loosening Characteristics of Self-Tapped and Non-Self-Tapped Cortical Bone ScrewsClinical Orthopaedics and Related Research, 1981
- Recovery of Diaphyseal Bone Strength After Rigid Internal Plate Fixation:An Experimental Study in the RabbitActa Orthopaedica, 1980
- Bone ingrowth and stress shielding with a porous surface coated fracture fixation plateJournal of Biomedical Materials Research, 1979
- The Healing of Experimental Fractures by Compression Osteosynthesis: I.Torsional StrengthActa Orthopaedica, 1979
- Force measurements in screw fixationJournal of Biomechanics, 1976
- The Effects of Rigidity of Internal Fixation Plates on Long Bone Remodeling:A Bio Mechanical and Quantitative Histological StudyActa Orthopaedica, 1976