Kinetic study of the Jaffé reaction for quantifying creatinine in serum: 2. Evaluation of buffered reagent and comparison of different data-processing options.
Open Access
- 1 March 1989
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 35 (3) , 360-363
- https://doi.org/10.1093/clinchem/35.3.360
Abstract
Here we describe the evaluation of several data-processing options for the kinetic determination of creatinine by use of the Jaffé reaction. Data-processing options evaluated include initial-rate, two-point fixed-time, rate at t = k-1, and multipoint curve-fitting predictive methods. We evaluated these options for a buffered formulation of the Jaffé reagent and studied the effects of potential interferents, including glucose, acetoacetate, bilirubin, and albumin, on each option. To reduce effects of bilirubin, we evaluated the inclusion of a preoxidation step with ferricyanide. All the data-processing options gave good precision and linearity between the measurement objective and creatinine concentration. However, differences between slopes of calibration plots in aqueous and serum matrices ranged from a high of +60% for the two-point, fixed-time method to a low of -11% for the curve-fitting, predictive method. Standard additions of creatinine to sera were quantified reliably (yielding 96% to 102% of target values) by the predictive method and less reliably (62% to 102%) by the other methods. We conclude that the predictive method has the potential to yield the most reliable results for creatinine.This publication has 7 references indexed in Scilit:
- Kinetic study of the Jaffé reaction for quantifying creatinine in serum: 1. Alkalinity controlled with NaOH.Clinical Chemistry, 1987
- Comparison of reaction-rate methods of analysis for systems following first-order kineticsAnalytical Chemistry, 1986
- Direct ultraviolet method for enzymatic determination of uric acid, with equilibrium and kinetic data-processing options.Clinical Chemistry, 1986
- KINETIC SERUM CREATININE ASSAYS .1. THE ROLE OF VARIOUS FACTORS IN DETERMINING SPECIFICITY1980
- KINETIC SERUM CREATININE ASSAYS .2. A CRITICAL-EVALUATION AND REVIEW1980
- A comprehensive classification of kinetic methods of analysis used in clinical chemistry.Clinical Chemistry, 1977
- Kinetic study of the cyanogen bromide-pyridine reaction applied to thiocyanate determinations in serumAnalytical Chemistry, 1977