Comparing Methods of Measurement for Detecting Drug‐Induced Changes in the QT Interval: Implications for Thoroughly Conducted ECG Studies
- 13 April 2004
- journal article
- clinical trial
- Published by Wiley in Annals of Noninvasive Electrocardiology
- Vol. 9 (2) , 166-174
- https://doi.org/10.1111/j.1542-474x.2004.92542.x
Abstract
The aim of this study was to compare the reproducibility and sensitivity of four commonly used methods for QT interval assessment when applied to ECG data obtained after infusion of ibutilide. Four methods were compared: (1) 12-lead simultaneous ECG (12-SIM), (2) lead II ECG (LEAD II), both measured on a digitizing board, (3) 3-LEAD ECG using a manual tangential method, and (4) a computer-based, proprietary algorithm, 12SL trade mark ECG Analysis software (AUT). QT intervals were measured in 10 healthy volunteers at multiple time points during 24 hours at baseline and after single intravenous doses of ibutilide 0.25 and 0.5 mg. Changes in QT interval from baseline were calculated and compared across ECG methods, using Bland-Altman plots. Variability was studied using a mixed linear model. Baseline QT values differed between methods (range 376-395 ms), mainly based on the number of leads incorporated into the measurement, with LEAD II and 3-LEAD providing the shortest intervals. The 3-LEAD generated the largest QT change from baseline, whereas LEAD II and 12-SIM generated essentially identical result within narrow limits of agreement (0.4 ms mean difference, 95% confidence interval +/- 20.5 ms). Variability with AUT (standard deviation 15.8 ms for within-subject values) was clearly larger than with 3-LEAD, LEAD II, and 12-SIM (9.6, 10.0, and 11.3 ms). This study demonstrated significant differences among four commonly used methods for QT interval measurement after pharmacological prolongation of cardiac repolarization. Observed large differences in variability of measurements will have a substantial impact on the sample size required to detect QT prolongation in the range that is currently advised in regulatory guidance.Keywords
This publication has 30 references indexed in Scilit:
- Focus on issues in measuring and interpreting changes in the QTc interval durationEuropean Heart Journal Supplements, 2001
- Spectrum of drugs prolonging QT interval and the incidence of torsades de pointesEuropean Heart Journal Supplements, 2001
- Evaluation of Drug-Induced QT Interval ProlongationDrug Safety, 2001
- Electrophysiology and Pharmacology of IbutilideThe American Journal of Cardiology, 1996
- Efficacy of intravenous ibutilide for rapid termination of atrial fibrillation and atrial flutter: A dose-response studyJournal of the American College of Cardiology, 1996
- Dose-response relation between terfenadine (Seldane) and the QTc interval on the scalar electrocardiogram: Distinguishing a drug effect from spontaneous variabilityAmerican Heart Journal, 1996
- Variability of the QTc interval: Impact on defining drug effect and low-frequency cardiac eventThe American Journal of Cardiology, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Prolongation of the Q-T interval in man during sleepThe American Journal of Cardiology, 1983
- Die Systolendauer im Elektrokardiogramm bei normalen Menschen und bei HerzkrankenActa Medica Scandinavica, 1920