Artefakte durch Dentallegierungen in der MR-Bildgebung des Temporomandibulargelenks: eine Phantommodellstudie bei 1,5 T
- 1 May 1997
- journal article
- research article
- Published by Georg Thieme Verlag KG in RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren
- Vol. 166 (05) , 421-428
- https://doi.org/10.1055/s-2007-1015452
Abstract
The influence of dental alloys on MRI of the temporomandibular joint was studied using a phantom model for this joint. At 1,5 T, 15 dental alloys and 14 of their most important components were investigated acquiring sagittal (FOV: 150 mm) and transverse (FOV: 250 mm) T1-weighted SE images. In 11 cases, T1- and T2*-weighted FLASH images were measured additionally. The artifacts were assessed qualitatively as well as quantitatively, and the samples were subdivided into four artifact categories. Ag, Cu, Ga, In, Tl, Sn, Zn, amalgam, the precious alloys, the Au-Pd and Ag-Pd alloys showed no artifacts (category I). Minimal artifacts below 10 mm on transverse images (category II) were found for Cr, Pd, Pt and for the Ni-Cr alloy. Mn and the remaining non-precious alloys induced artifacts up to 30 mm (category III). Significant artifacts-more than 30 mm-(category IV) were caused by Ni-Cr and 18/8 wires and by Co, Fe, and Ni. T2*-weighted FLASH proved to be more susceptible for artifacts than T1-weighted SE and FLASH techniques. In contrast to dental alloys for fixed prosthodontics, Ni-Cr- or 18/8 wires used for orthodontic bands can influence not only the image quality, but also the diagnostic reliability of MRI of the temporomandibular joint.Keywords