The physical and bacteriological properties of disposable and non-disposable suction drainage units in the laboratory
- 1 October 1974
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Surgery
- Vol. 61 (10) , 832-837
- https://doi.org/10.1002/bjs.1800611020
Abstract
The physical and bacteriological properties of disposable and non-disposable suction units have been studied in the laboratory. The disposable units which operate at a lower suction pressure were found to be subject to spontaneous loss of vacuum, while positional change of the containers of the magnitude encountered in clinical practice could easily produce reversal of flow from the container. The incidence of contamination in the systems in a laboratory model designed to simulate the ward situation was higher in the disposable units.Keywords
This publication has 6 references indexed in Scilit:
- A controlled trial of closed wound suction drainage in radical mastectomyBritish Journal of Surgery, 1973
- Suction drainage. A study in wound healingBritish Journal of Surgery, 1970
- HAZARD OF VACU-DRAIN SYSTEMThe Lancet, 1969
- Disappearing drain.BMJ, 1968
- THE USE OF NEGATIVE PRESSURE UNDER SKIN FLAPS AFTER RADICAL MASTECTOMYAnnals of Surgery, 1952
- DRAINAGE IN ABDOMINAL SURGERY.JAMA, 1890