Comparison of Two Methods for Selecting Minimum Stimulation Levels Used in Programming the Nucleus 22 Cochlear Implant
- 1 August 1999
- journal article
- research article
- Published by American Speech Language Hearing Association in Journal of Speech, Language, and Hearing Research
- Vol. 42 (4) , 814-828
- https://doi.org/10.1044/jslhr.4204.814
Abstract
Minimum stimulation levels for active electrodes in a Nucleus 22 cochlear implant were set at threshold (clinical default value) and raised levels ( M =+2.04 dB) to determine if raised levels would improve recipients' understanding of soft speech sounds with the SPEAK speech coding strategy. Eight postlinguistically deaf adults participated in a 4-phase A 1 B 1 A 2 B 2 , test design. Speech recognition was evaluated with consonant-vowel nucleus-consonant (CNC) words in quiet and sentences in noise, both presented at 50, 60, and 70 dB SPL during 2 weekly sessions at the end of each phase. Group mean scores were significantly higher with the raised level program for words and phonemes at 50 and 60 dB SPL and for sentences at 50 and 70 dB SPL. All participants chose to use the raised level program in everyday life at the end of the study. The results suggest that clinical use of a raised level program for Nucleus 22 recipients has the potential to make soft sounds louder and, therefore, more salient in everyday life. Further research is needed to determine if this approach is appropriate for other cochlear implant devices.Keywords
This publication has 10 references indexed in Scilit:
- Parameter selection to optimize speech recognition with the Nucleus implantOtolaryngology -- Head and Neck Surgery, 1997
- Speech recognition at simulated soft, conversational, and raised-to-loud vocal efforts by adults with cochlear implantsThe Journal of the Acoustical Society of America, 1997
- Use of Test-Retest Measures to Evaluate Performance Stability in Adults with Cochlear ImplantsEar & Hearing, 1995
- Performance of Postlinguistically Deaf Adults with the Wearable Speech Processor (WSP III) and Mini Speech Processor (MSP) of the Nucleus Multi-Electrode Cochlear Implant*Ear & Hearing, 1991
- Voice Fundamental Frequency as an Auditory Supplement to the Speechreading of SentencesEar & Hearing, 1988
- Composite Speech Spectrum for Hearing Aid Gain PrescriptionsJournal of Speech, Language, and Hearing Research, 1988
- Sound Field AudiometryEar & Hearing, 1984
- AN EXPANDED TEST FOR SPEECH DISCRIMINATION UTILIZING CNC MONOSYLLABIC WORDS: NORTHWESTERN UNIVERSITY AUDITORY TEST NO. 6Published by Defense Technical Information Center (DTIC) ,1966
- Revised CNC Lists for Auditory TestsJournal of Speech and Hearing Disorders, 1962
- Preferred Method For Clinical Determination Of Pure-Tone ThresholdsJournal of Speech and Hearing Disorders, 1959