Bone mass measurement by DXA: Influence of analysis procedures and interunit variation
- 1 December 1992
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Bone and Mineral Research
- Vol. 7 (12) , 1373-1382
- https://doi.org/10.1002/jbmr.5650071204
Abstract
Interunit variability among bone densitometers is due to different factors, including different calibration procedures and algorithms and variability in photon source energies and/or intensities. Other factors, such as the choice of scan parameters or the analysis procedures, can also introduce variability. The new generation of dual-energy x-ray absorptiometry (DXA) has partially improved this situation. The aim of this study was to investigate the operator-dependent analysis procedures that can affect scan results and to evaluate the phantom and in vivo interunit variation of some DXA instruments. Four DXA instruments (QDR 1000 and 1000/W, Hologic, Inc.) were used. Potential sources of variability in the analysis procedures of anteroposterior lumbar spine and hip scans were considered: in most cases these procedures significantly influenced scan results. On lumbar spine, an enlargement of the scan window of less than 3 cm was responsible for an average increase in bone mineral density (BMD) of about 3%. On the hip, lowering the scan window by about 1 cm accounted for an increase in the whole-segment BMD of about 4%. After standardization of analysis procedures, interunit and intraunit coefficients of variation and percentage differences among instruments were less than 1% for all the parameters considered (area and bone mineral content and density) with both an anatomic and a geometric phantom, and in nine subjects scanned by two different devices the percentage difference in BMD was greater than 2%. This study shows that present interunit variability allows comparisons among laboratories, but only if highly standardized analysis procedures are used.Keywords
Funding Information
- CNR (91.00420.PF40)
This publication has 11 references indexed in Scilit:
- The effect of positioning on dual energy X-ray bone densitometry of the proximal femurBone and Mineral, 1991
- The clinical application of serial bone mass measurementsBone and Mineral, 1991
- Spinal and forearm bone mass in relation to ageing and menopause in healthy Italian womenEuropean Journal of Clinical Investigation, 1991
- Calibration and standardization of bone mineral densitometersJournal of Bone and Mineral Research, 1989
- Bone Densitometry for Diagnosis and Monitoring OsteoporosisExperimental Biology and Medicine, 1989
- Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar SpineMayo Clinic Proceedings, 1988
- Precision error in dual-photon absorptiometry related to source age.Radiology, 1988
- Dual photon absorptimetric measurements of bone mineral density increase with source lifeCalcified Tissue International, 1987
- Quantitative CT applications: comparison of current scanners.Radiology, 1987
- QUANTITATIVE COMPUTED TOMOGRAPHY FOR SPINAL MINERAL ASSESSMENTJournal of Computer Assisted Tomography, 1985