Comparison of the New OPS Imaging Technique with Intravital Microscopy: Analysis of the Colon Microcirculation
- 14 March 2002
- journal article
- research article
- Published by S. Karger AG in European Surgical Research
- Vol. 34 (1-2) , 124-128
- https://doi.org/10.1159/000048898
Abstract
Background: The OPS imaging technique has been introduced for in vivo assessment of microcirculation in humans. The aim of this study was to validate the new technique against intravital fluorescence microscopy (IFM) for the visualization of colon microcirculation in a murine model of inflammatory bowel disease (IBD). Method: IBD was induced in Balb/c mice by dextran sulfate sodium, controls received normal water. In each animal, both the CYTOSCANTM A/R and IFM were used to image the microcirculation (n = 7 in each group). The postcapillary venular diameter was analyzed on the colon muscularis and mucosa. Results: The venular diameter correlated significantly between both methods representing the good correspondence between both methods. Conclusion: Our study demonstrates that the new technique for visualization of microcirculation without use of fluorescent dyes, the OPS imaging, allows for quantitative measurement of a key microcirculatory parameters of the mouse colon.Keywords
This publication has 3 references indexed in Scilit:
- The CytoscanTM Model E-II, a New Reflectance Microscope for Intravital Microscopy: Comparison with the Standard Fluorescence MethodJournal of Vascular Research, 2000
- Orthogonal polarization spectral imaging: A new method for study of the microcirculationNature Medicine, 1999
- Quantitative analysis of small intestinal microcirculation in the mouseZeitschrift für Die Gesamte Experimentelle Medizin, 1998