Surface flatness after bone cutting: A cadaver study of tibial condyles
- 1 January 1991
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 62 (1) , 15-18
- https://doi.org/10.3109/17453679108993084
Abstract
A methodologic study with which the cut surface could be quantified was performed on cadaveric tibial bone prepared for endoprosthetic components. Using sterilized, dental-imprint material and measuring with a Zeiss UMC 850, the characteristics of the cut surface were defined. A clinically flat surface was found to be uneven, with a maximum roughness between the uppermost and lowermost points of 1.0 and 2.4 mm for the bone surface and between 1.2 and 2.3 mm for the imprint. The flatness, defined as the standard deviation of the measuring points, was between 0.15 and 0.40 mm for the bone and 0.20 and 0.42 mm for the imprint. This lack of flatness creates gaps between the prosthesis and the bone.Keywords
This publication has 2 references indexed in Scilit:
- Implant fixation improved by close fit Cylindrical implant – bone interface studied in rabbitsActa Orthopaedica, 1988
- Biological fixation of porous-coated implantsJournal of Bone and Joint Surgery, 1987