Water calorimeter dosimetry for 160 MeV protons
- 1 April 1992
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 37 (4) , 947-953
- https://doi.org/10.1088/0031-9155/37/4/009
Abstract
The absorbed dose to water from a 160 MeV proton beam as determined by a flexible, temperature regulated, sealed glass core, water calorimeter was compared to that determined from ionization chambers used in accordance with AAPM Report 16. The ratios of these doses as obtained from two experiments done over four months apart, are 0.992+or-0.004 and 0.990+or-0.004. As there are no radiation dependent parameters required for the water calorimeter, these data add to the growing body of evidence which supports the use of the calorimeter as a reliable absorbed dose standard. They also support the use of 60Co-calibrated ionization chambers used in accordance with AAPM Report 16 for the dosimetry of proton beams.Keywords
This publication has 9 references indexed in Scilit:
- A comparison of ionization‐chamber and water‐calorimeter dosimetry for high‐energy x raysMedical Physics, 1991
- The thermal defect of A-150 plastic and graphite for low-energy protonsPhysics in Medicine & Biology, 1990
- A direct comparison of water calorimetry and Fricke dosimetryPhysics in Medicine & Biology, 1989
- The direct determination of dose‐to‐water using a water calorimeterMedical Physics, 1987
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983
- Measurements of the thermal defect for A-150 plastic (tissue-equivalent plastic)Physics in Medicine & Biology, 1982
- Flattening of proton dose distributions for large‐field radiotherapyMedical Physics, 1977
- Range modulators for protons and heavy ionsNuclear Instruments and Methods, 1975
- Measurements of heat capacity and heat of vaporization of water in the range 0 degrees to 100 degrees CJournal of Research of the National Bureau of Standards, 1939