Robustness and precision of an automatic marker detection algorithm for online prostate daily targeting using a standard V‐EPID
- 25 June 2003
- journal article
- clinical trial
- Published by Wiley in Medical Physics
- Vol. 30 (7) , 1825-1832
- https://doi.org/10.1118/1.1584041
Abstract
An algorithm for the daily localization of the prostate using implanted markers and a standard video‐based electronic portal imaging device (V‐EPID) has been tested. Prior to planning, three gold markers were implanted in the prostate of seven patients. The clinical images were acquired with a2.1 V‐EPID for each field during the normal course radiotherapy treatment and are used off‐line to determine the ability of the automatic marker detection algorithm to adequately and consistently detect the markers. Clinical images were obtained with various dose levels from ranging 2.5 to 75 MU. The algorithm is based on marker attenuation characterization in the portal image and spatial distribution. A total of 1182 clinical images were taken. The results show an average efficiency of 93% for the markers detected individually and 85% for the group of markers. This algorithm accomplishes the detection and validation in 0.20–0.40 s. When the center of mass of the group of implanted markers is used, then all displacements can be corrected to within 1.0 mm in 84% of the cases and within 1.5 mm in 97% of cases. The standard video‐based EPID tested provides excellent marker detection capability even with low dose levels. The V‐EPID can be used successfully with radiopaque markers and the automatic detection algorithm to track and correct the daily setup deviations due to organ motions.Keywords
This publication has 24 references indexed in Scilit:
- Clinical use of electronic portal imaging: Report of AAPM Radiation Therapy Committee Task Group 58Medical Physics, 2001
- A quality control test for electronic portal imaging devicesMedical Physics, 1996
- Evaluation of changes in the size and location of the prostate, seminal vesicles, bladder, and rectum during a course of external beam radiation therapyInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Automated localization of the prostate at the time of treatment using implanted radiopaque markers: Technical feasibilityInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Prostate motion during standard radiotherapy as assessed by fiducial markersRadiotherapy and Oncology, 1995
- Measurement of prostate movement over the course of routine radiotherapy using implanted markersInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Localization of the prostatic apex for radiation therapy using implanted markersInternational Journal of Radiation Oncology*Biology*Physics, 1993
- Automated determination of patient setup errors in radiation therapy using spherical radio‐opaque markersMedical Physics, 1993
- Treatment planning issues related to prostate movement in response to differential filling of the rectum and bladderInternational Journal of Radiation Oncology*Biology*Physics, 1991
- Adaptive histogram equalization and its variationsComputer Vision, Graphics, and Image Processing, 1987