Turbulent Stresses in the Region of a Hancock Porcine Bioprosthetic Aortic Valve
- 1 August 1985
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 107 (3) , 200-205
- https://doi.org/10.1115/1.3138544
Abstract
The purpose of this study was to measure stresses associated with turbulence (Reynolds stresses), in the region of a 29-mm-dia porcine bioprosthetic valve (Hancock, Model 242). Studies were performed in an in vitro pulse duplicating system with the valve mounted in the aortic position. The Reynolds stresses were calculated from velocities obtained with a two channel laser Doppler anemometer. The largest Reynolds shear stress and normal stress occurred at the highest stroke volume used (80 mL). Averaged over ejection they were 38 dynes/cm2 and 380 dynes/cm2 , respectively. The maximal instantaneous Reynolds shear stress was 2500 dynes/cm2 and the maximal instantaneous Reynolds normal stress was 6800 dynes/cm2 . Stresses of these magnitudes are in the range reported to damage platelets.Keywords
This publication has 8 references indexed in Scilit:
- Wall shear stress during pulsatile flow distal to a normal porcine aortic valveJournal of Biomechanics, 1984
- Platelet reactivity in patients with degenerated porcine bioprosthetic valvesThrombosis Research, 1981
- Surface morphology of degenerated porcine bioprosthetic valves four to seven years following implantationThe Journal of Thoracic and Cardiovascular Surgery, 1981
- Contribution of semilunar leaflets to turbulent blood flowBiorheology, 1979
- In vitro velocity measurements in the vicinity of aortic prosthesesJournal of Biomechanics, 1979
- Input impedance of the systemic circulation in man.Circulation Research, 1977
- Turbulent blood flow in the ascending aorta of humans with normal and diseased aortic valves.Circulation Research, 1976
- Use of Random Excitation and Spectral Analysis in the Study of Frequency-Dependent Parameters of the Cardiovascular SystemCirculation Research, 1966