Neurotransmitter release from high-frequency stimulation of the subthalamic nucleus
- 1 September 2004
- journal article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 101 (3) , 511-517
- https://doi.org/10.3171/jns.2004.101.3.0511
Abstract
High-frequency stimulation (HFS) delivered through implanted electrodes in the subthalamic nucleus (STN) has become an established treatment for Parkinson disease (PD). The precise mechanism of action of deep brain stimulation (DBS) in the STN is unknown, however. In the present study, the authors tested the hypothesis that HFS within the STN changes neuronal action potential firing rates during the stimulation period by modifying neurotransmitter release. Intracellular electrophysiological recordings were obtained using sharp electrodes in rat STN neurons in an in vitro slice preparation. A concentric bipolar stimulating electrode was placed in the STN slice, and electrical stimulation (pulse width 50-100 microsec, duration 100-2000 msec, amplitude 10-500 microA, and frequency 10-200 Hz) was delivered while simultaneously obtaining intracellular recordings from an STN neuron. High-frequency stimulation of the STN either generated excitatory postsynaptic potentials (EPSPs) and increased the action potential frequency or it generated inhibitory postsynaptic potentials and decreased the action potential frequency of neurons within the STN. These effects were blocked after antagonists to glutamate and gamma-aminobutyric acid were applied to the tissue slice, indicating that HFS resulted in the release of neurotransmitters. Intracellular recordings from substantia nigra pars compacta (SNc) dopaminergic neurons during HFS of the STN revealed increased generation of EPSPs and increased frequency of action potentials in SNc neurons. During HFS of STN neurons the mechanism of DBS may involve the release of neurotransmitters rather than the primary electrogenic inhibition of neurons.Keywords
This publication has 50 references indexed in Scilit:
- Stimulation of the Subthalamic Nucleus in Parkinson's Disease Does Not Produce Striatal Dopamine ReleaseNeurosurgery, 2003
- High-Frequency Stimulation Produces a Transient Blockade of Voltage-Gated Currents in Subthalamic NeuronsJournal of Neurophysiology, 2001
- High frequency stimulation of the STN influences the activity of dopamine neurons in the ratNeuroReport, 2000
- Modulation of dopamine efflux in the nucleus accumbens after cholinergic stimulation of the ventral tegmental area in intact, pedunculopontine tegmental nucleus-lesioned, and laterodorsal tegmental nucleus-lesioned ratsJournal of Neuroscience, 1996
- Responses of substantia nigra pars reticulata and globus pallidus complex to high frequency stimulation of the subthalamic nucleus in rats: electrophysiological dataNeuroscience Letters, 1995
- Acute and Long-Term Effects of Subthalamic Nucleus Stimulation in Parkinson's DiseaseStereotactic and Functional Neurosurgery, 1994
- Reversal of Rigidity and Improvement in Motor Performance by Subthalamic High‐frequency Stimulation in MPTP‐treated MonkeysEuropean Journal of Neuroscience, 1993
- Application of in vivo electrochemistry to the measurement of changes in dopamine release during intracranial self-stimulationJournal of Neuroscience Methods, 1990
- Intracellular studies in the facial nucleus illustrating a simple new method for obtaining viable motoneurons in adult rat brain slicesSynapse, 1989