Comparative Validation of Quantitative Coronary Angiography Systems
- 15 April 1995
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 91 (8) , 2174-2183
- https://doi.org/10.1161/01.cir.91.8.2174
Abstract
Background Computerized quantitative coronary angiography (QCA) has fundamentally altered our approach to the assessment of coronary interventional techniques and strategies aimed at the prevention of recurrence and progression of stenosis. It is essential, therefore, that the performance of QCA systems, upon which much of our scientific understanding has become integrally dependent, is evaluated in an objective and uniform manner. Methods and Results We validated 10 QCA systems at core laboratories in North America and Europe. Cine films were made of phantom stenoses of known diameter (0.5 to 1.9 mm) under four experimental conditions: in vivo (coronary arteries of pigs) calibrated at the isocenter or by use of the catheter as a scaling device and in vitro with 50% contrast and 100% contrast. The cine films were analyzed by each automated QCA system without observer interaction. Accuracy and precision were taken as the mean and SD of the signed differences between the phantom stenoses, and the measured minimal luminal diameters and the correlation coefficient ( r ), the SEE, the y intercept, and the slope were derived by their linear regression. Performance of the 10 QCA systems ranged widely: accuracy, +0.07 to +0.31 mm; precision, ±0.14 to ±0.24 mm; correlation ( r ), .96 to .89; SEE, ±0.11 to ±0.16 mm; intercept, +0.08 to +0.31 mm; and slope, 0.86 to 0.64. Conclusions There is a marked variability in performance between systems when assessed over the range of 0.5 to 1.9 mm. The range of accuracy, intercept, and slope values of this report indicates that absolute measurements of luminal diameter from different multicenter angiographic trials may not be directly comparable and additionally suggests that such absolute measurements may not be directly applicable to clinical practice using an on-line QCA system with a different edge detection algorithm. Power calculations and study design of angiographic trials should be adjusted for the precision of the QCA system used to avoid the risk of failing to detect small differences in patient populations. This study may guide the fine-tuning of algorithms incorporated within each system and facilitate the maintenance of high standards of QCA for scientific studies.Keywords
This publication has 43 references indexed in Scilit:
- A new approach for the quantification of complex lesion morphology: The gradient field transform; Basic principles and validation resultsJournal of the American College of Cardiology, 1994
- Characteristics and radiofrequency ablation therapy of intermediate septal accessory pathwayThe American Journal of Cardiology, 1994
- Usefulness of repeat coronary angiography 24 hours after successful balloon angioplasty to evaluate early luminal deterioration and facilitate quantitative analysisThe American Journal of Cardiology, 1993
- Restenosis revisited: Insights provided by quantitative coronary angiographyAmerican Heart Journal, 1993
- Can the same edge-detection algorithm be applied to on-line and off-line analysis systems? Validation of a new cinefilm-based geometric coronary measurement softwareAmerican Heart Journal, 1993
- Quantitative analysis of factors influencing late lumen loss and restenosis after directional coronary atherectomyThe American Journal of Cardiology, 1993
- Calibration using angiographic catheters as scaling devices — Importance of filming the catheters not filled with contrast mediumThe American Journal of Cardiology, 1992
- Comparison of quantitative coronary angiography to visual estimates of lesion severity pre and post PTCAAmerican Heart Journal, 1990
- Comparison of coronary stenosis quantitation results from on-line digital and digitized cine film imagesThe American Journal of Cardiology, 1988
- Change in diameter of coronary artery segments adjacent to stenosis after percutaneous transluminal coronary angioplasty: Failure of percent diameter stenosis measurement to reflect morphologic changes induced by balloon dilationJournal of the American College of Cardiology, 1988