A Hyperthermia System Using a New Type of Inductive Applicator
- 1 January 1984
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. BME-31 (1) , 21-27
- https://doi.org/10.1109/tbme.1984.325366
Abstract
A hyperthermia system working at frequencies around 150 MHz is described. It is useful for a penetration depth of 3-4 cm in tumor and muscle tissue. The system includes a new type of inductive applicator giving a beam-like distribution in the tissue, a controllable surface cooling system, and multipoint temperature measurements. A phase I clinical study shows the efficacy of the system, but also the importance of temperature measurements at many points since, in general, the temperature gradients are large.Keywords
This publication has 5 references indexed in Scilit:
- Near‐field absorption in a circular cylinder from electric and magnetic line sourcesBioelectromagnetics, 1982
- Fractionated radiation and hyperthermia: Experimental and clinical studiesCancer, 1981
- Electronics in medicine: RF therapy for malignancy: Heating of malignant tissues (hyperthermia) by RF radiation presents a new tool in the arsenal of weapons against cancerIEEE Spectrum, 1980
- Therapeutic applications of electromagnetic powerProceedings of the IEEE, 1974
- Nonionizing electromagnetic wave effects in biological materials and systemsProceedings of the IEEE, 1972