Bragg-reflection profiles of graphite and alkali–graphite intercalation compounds: Comparison of double-axis and triple-axis spectrometer results
- 1 February 1983
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 54 (2) , 206-209
- https://doi.org/10.1063/1.1137346
Abstract
Double-axis x-ray spectrometer measurements at flat, low-absorbing mosaic crystals produce broad and asymmetric Bragg-reflection profiles. If the absorption coefficient changes during the sample preparation process, as in the case of graphite and graphite intercalation compounds, a determination of the instrumental profile is almost impossible. Using triple-axis spectrometers, the Bragg-reflection profiles are independent of the sample transparency, and peak profiles can be determined with high resolution and minor corrections.Keywords
This publication has 4 references indexed in Scilit:
- Inelastic neutron measurement of phonons in graphite-alkali intercalation compoundsPhysical Review B, 1982
- Intercalation compounds of graphiteAdvances in Physics, 1981
- The Effect of a Low Absorption Coefficient on X-Ray Spectrometer MeasurementsReview of Scientific Instruments, 1952
- A Numerical Fourier-analysis Method for the Correction of Widths and Shapes of Lines on X-ray Powder PhotographsProceedings of the Physical Society, 1948