Okulo-Oszillo-Dynamographie nach Ulrich: Erste Ergebnisse bei Carotisstenosen
- 1 October 1985
- journal article
- research article
- Published by Georg Thieme Verlag KG in Klinische Monatsblätter für Augenheilkunde
- Vol. 187 (10) , 256-261
- https://doi.org/10.1055/s-2008-1051029
Abstract
Oculo-oscillodynamography (OODG) after Ulrich was perfored in 11 patients suffering from a monolateral, hemodynamically relevant carotid stenosis identified by Doppler sonography. All of the pressure parameters which can be measured by OODG-systolic retinal, systolic ciliary, and diastolic ocular perfusion pressure - were low on the side concerned. In contrast to the already-established methods of measuring ocular perfusion-ophthalmodynamometry and ophthalmodynamography after Hager-the advantages of OODG are that it not only enables the systolic retinal pressure to be differentiated from the systolic ciliary or diastolic ocular perfusion pressure, but also permits the results of the investigation to be illustrated graphically, so that objective evaluation is possible.This publication has 2 references indexed in Scilit:
- Acute Central Retinal Artery Occlusion and Optic Disc NeovascularizationArchives of Ophthalmology (1950), 1978
- Experiences with Ophthalmodynamography in the Diagnosis of Carotid OcclusionAmerican Journal of Ophthalmology, 1967