The Causes of Femoral Head Roughening in Explanted Charnley Hip Prostheses
- 1 July 1987
- journal article
- research article
- Published by SAGE Publications in Engineering in Medicine
- Vol. 16 (3) , 167-173
- https://doi.org/10.1243/emed_jour_1987_016_035_02
Abstract
The effect of particles of first, acrylic bone cement containing either barium sulphate or zirconium dioxide as X-ray contrast media, and secondly, particles of fresh cortical bone, on a surgical grade stainless steel surface against which polyethylene was rubbing, has been investigated. It was found that under physiological loads, particles of acrylic cement which have been embedded in the polyethylene bearing surface caused severe scratching of the stainless steel. Some scratches were found to have pronounced lips similar to those observed on the femoral heads of explanted Charnley hip prostheses. Cortical bone produced only slight damage to the stainless steel. In the case of the acrylic cements, the abrasive constituent was isolated and found to be clusters of particles of the X-ray contrast medium. Particles of acrylic cement are frequently observed embedded in the wear surfaces of explanated Charnley acetabular cups and these particles have also been found to contain clusters of X-ray contrast medium particles. It is therefore concluded that the damage to implanted stainless steel prostheses is predominantly caused by the X-ray contrast medium in the acrylic cement.This publication has 10 references indexed in Scilit:
- The Role of Cement in the Long Term Performance and Premature Failure of Charnley Low Friction ArthroplastiesEngineering in Medicine, 1986
- Walking activities and wear of prostheses.Annals of the Rheumatic Diseases, 1985
- Direction and rate of socket wear in Charnley low-friction arthroplastyThe Journal of Bone and Joint Surgery. British volume, 1985
- Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints: III: The measurement of internal volume changes in explanted Charnley sockets after 2–16 years in vivo and the determination of wear factorsWear, 1985
- Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints: II: A microscopical study of the surfaces of Charnley polyethylene acetabular socketsWear, 1985
- The appearances of wear on polyethylene—a comparison of in vivo and in vitro wear surfacesJournal of Biomedical Materials Research, 1978
- The state of the art in joint replacementJournal of Medical Engineering & Technology, 1978
- Reactions of the articular capsule to wear products of artificial joint prosthesesJournal of Biomedical Materials Research, 1977
- Elements of hip joint prosthesis reliabilityJournal of Medical Engineering & Technology, 1977
- Clinical problems in total hip replacementsJournal of Biomedical Materials Research, 1974