Oxygen electrode design criteria and performance characteristics: recessed cathode
- 1 July 1978
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 45 (1) , 145-154
- https://doi.org/10.1152/jappl.1978.45.1.145
Abstract
A computer simulation of the steady-state operation of recessed (Whalen-type) polarographic oxygen electrodes has been developed to give the design factors important for performance optimization. The simulation makes use of a specially formulated three-dimensional orthogonal coordinate system with the geometry identical to the actual recessed cathode and gives the oxygen concentration field induced by it in the surrounding medium. Equations are presented which allow one to calculate, for any recessed cathode, the current sensitivity, maximum stirring artifact, measurement error, and time constant. Comparisons with analytically obtained expressions for the corresponding quantities for idealized, spherosymmetric cathodes demonstrate the unique aspects of recessed-cathode performance. For commonly used electrodes, a recess length-to-cathode diameter ratio of greater than 10 is found to give a negligible stirring artifact, a negligible measurement error, and a rapid response.This publication has 5 references indexed in Scilit:
- A microelectrode for measuring intracellular PO2.Journal of Applied Physiology, 1967
- Measurement of oxygen tension by the oxygen cathodeMedical & Biological Engineering & Computing, 1967
- An ultramicro oxygen electrodeJournal of Applied Physiology, 1964
- PROBLEMS ASSOCIATED WITH THE USE OF THE EXPOSED PLATINUM ELECTRODE FOR MEASURING OXYGEN TENSION IN VIVOCanadian Journal of Biochemistry and Physiology, 1958
- Microelectrodes for Measuring Local Oxygen Tension in Animal TissuesReview of Scientific Instruments, 1942