Compliance properties of conduits used in vascular reconstruction
Open Access
- 1 November 2000
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Surgery
- Vol. 87 (11) , 1516-1524
- https://doi.org/10.1046/j.1365-2168.2000.01566.x
Abstract
Background: Compliance mismatch between native artery and prosthetic graft used for infrainguinal bypass is implicated in the aetiology of graft failure. The aim was to quantify the elastic properties of a new compliant poly(carbonate)polyurethane (CPU) vascular graft, and to compare the compliance properties of grafts made from CPU, expanded polytetrafluoroethylene (ePTFE), Dacron and human saphenous vein with that of human muscular artery. Methods: A pulsatile flow phantom was used to perfuse vessel and prosthetic graft segments at physiological pulse pressure and flow. Intraluminal pressure was measured using a Millar Mikro-tip catheter transducer and vessel wall motion was determined with duplex ultrasonography using an echo-locked wall-tracking system. Diametrical compliance and a stiffness index were then calculated for each type of conduit over mean pressures ranging from 30 to 100 mmHg by 10-mmHg increments. Results: The compliance values of CPU and artery (mean over the pressure range) were similar (mean(s.d.) 8·1(0·4) and 8·0(5·9) per cent per mmHg × 10−2 respectively), although the elastic behaviour of artery was anisotropic unlike CPU, which was isotropic. Dacron and ePTFE grafts had lower compliance values (1·8(1·2) and 1·2(0·3) per cent per mmHg × 10−2 respectively, averaged over the pressure range). In both these cases, compliance and stiffness differed significantly from that of artery over a mean pressure range of 30–90 mmHg. Human saphenous vein exhibited anisotropic behaviour and, although compliant at low pressure (30 mmHg), was markedly incompliant at higher pressures. Conclusion: Compliant polyurethane grafts offer a greater degree of compliance match than either ePTFE or Dacron.Keywords
This publication has 28 references indexed in Scilit:
- Arterial Wall Mechanics in Conscious DogsCirculation Research, 1995
- Evaluation of reproducibility of a vessel wall movement detector system for assessment of large artery propertiesCardiovascular Research, 1994
- Non‐invasive measurement of pulsatile vessel diameter change and elastic properties in human arteries: a methodological study,Clinical Physiology and Functional Imaging, 1993
- In vivo Testing of a Biostable PolyurethaneJournal of Biomaterials Applications, 1991
- Assessment of the distensibility of superficial arteriesUltrasound in Medicine & Biology, 1990
- Compliance Effect on Patency of Small Diameter Vascular GraftsAnnual Northeast Bioengineering Conference, 1989
- Non-invasive assessment of the age related changes in stiffness of major branches of the human arteriesCardiovascular Research, 1987
- Biodegradation of a polyurethane in vitroJournal of Biomedical Materials Research, 1987
- Morphologic alteration of cultured arterial smooth muscle cells by cyclic stretchingJournal of Surgical Research, 1983
- Mechanical Properties of Arteries in VivoCirculation Research, 1960