An Analysis of the Performance of a Variable Venturi-type Oxygen Mask
Open Access
- 1 February 1980
- journal article
- research article
- Published by SAGE Publications in Anaesthesia and Intensive Care
- Vol. 8 (1) , 44-51
- https://doi.org/10.1177/0310057x8000800109
Abstract
The theoretical performance of the Hudson Multivent mask is considered. A method is described of determining the flow-weighted mean inspired oxygen concentration produced by the mask. Using this method, it was found that the inspired oxygen concentration was predictable only at the 24% setting using the recommended flow rates, but that increasing the supplied-oxygen flow rates above the recommended levels resulted in the return of predictable function. At settings above 30% predictable function was found to be unlikely, whatever the supplied oxygen flow rate. Recommendations are made regarding the use of this device.This publication has 12 references indexed in Scilit:
- Oxygen Concentrations Received From Commonly Used Delivery SystemsSouthern Medical Journal, 1978
- CONTROLLED OXYGEN THERAPY AT 60% CONCENTRATIONThe Lancet, 1976
- Actual Tracheal Oxygen Concentrations with Commonly Used Oxygen EquipmentAnesthesiology, 1976
- Measurement of intratracheal oxygen concentrations during face mask administration of oxygen: A modification for improved controlCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1975
- Oxygen TherapyPublished by American Medical Association (AMA) ,1974
- Oxygen Concentration Delivered by Venturi Masks With In-Line HumidificationJAMA, 1971
- Evaluation of Oxygen Therapy Devices by Arterial Oxygen TensionsDiseases of the Chest, 1969
- Comparative Evaluation of Oxygen Therapy TechniquesJAMA, 1962
- OXYGEN THERAPYArchives of Surgery, 1931