Characteristics of the absorbed dose to water standard at ENEA
- 1 April 1996
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 41 (4) , 657-674
- https://doi.org/10.1088/0031-9155/41/4/006
Abstract
The primary standard of absorbed dose to water established at ENEA for the Co-60 gamma-ray quality is based on a graphite calorimeter and an ionometric transfer system. This standard was recently improved after a more accurate assessment of some perturbation effects in the calorimeter and a modification of the water phantom shape and size. The conversion procedure requires two corresponding depths, one in graphite and one in water, where the radiation energy spectra must be the same. The energy spectra at the corresponding points were determined by a Monte Carlo simulation in water and graphite scaled phantoms. A thorough study of the calorimeter gap effect corrections was also made with regard to their dependence on depth and field size. A comparison between the ionization chamber calibration procedures based on the standards of absorbed dose to water and of air kerma was also made, confirming the consistency of the two methods.Keywords
This publication has 18 references indexed in Scilit:
- Experimental determination of the beam quality dependence factors, kQ, for ionization chambers used in photon and electron dosimetryPhysics in Medicine & Biology, 1995
- Absorbed-dose calibrations in high-energy photon beams at the National Physical Laboratory: conversion procedurePhysics in Medicine & Biology, 1994
- Absorbed dose beam quality factors for the dosimetry of high-energy photon beamsPhysics in Medicine & Biology, 1992
- Uncertainties in dosimetric data and beam calibrationInternational Journal of Radiation Oncology*Biology*Physics, 1990
- A proposal for the calibration of plane‐parallel ion chambers by accredited dosimetry calibration laboratoriesMedical Physics, 1990
- Gap correction for the calorimetric measurement of absorbed dose in graphite with a60Co beamPhysics in Medicine & Biology, 1989
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983
- Displacement corrections used in absorbed dose determinationMedical Physics, 1980
- The Partition of Kerma to Account For BremsstrahlungHealth Physics, 1979