Accuracy and reliability testing of a portable soft tissue indentor
- 1 June 2001
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Vol. 9 (2) , 232-240
- https://doi.org/10.1109/7333.928583
Abstract
We have designed, built, and tested a portable indentor device that allows us to determine force/displacement (F/D) measurements on soft tissue in a clinical or research setting. The indentor system consists of a load cell mounted on a three-dimensional measurement device (Metrecom). The output of the load cell and the Metrecom are recorded and analyzed by software running on a notebook computer. The displacement calibration of the Metrecom gave an average error=0.005 mm [standard deviation (SD)=0.062)]. The force calibration of the load cell resulted in an average error=0.022 N (SD=0.049) and a linearity of 1.0062(R/sup 2/=0.9998). The indentor device was tested on six different human soft tissues by two different investigators. The interreliabilities and intrareliabilities were 0.99 [interclass correlation (ICC)] indicating that the results were repeatable by more than one investigator. F/D measurements from indentor testing on two materials were comparable to values measured using an Instron device (5.34 versus, 5.52 N/mm, and 0.98 versus 1.04 N/mm). The device was used to measure the soft tissue characteristics on the plantar surface of the foot of one subject. These data were used to calculate the effective Young's modulus for the tissue using equations derived by Zheng et al. [1999] and indicated a wide range of values dependent upon the portion of the F/D curve used. All results indicate data from this portable indentor device are reliable, accurate, and sensitive enough to identify mechanical properties of human tissues.Keywords
This publication has 12 references indexed in Scilit:
- A mathematical analysis for indentation tests of articular cartilagePublished by Elsevier ,2004
- Effective elastic properties for lower limb soft tissues from manual indentation experimentIEEE Transactions on Rehabilitation Engineering, 1999
- In vivo indentation of lower extremity limb soft tissuesIEEE Transactions on Rehabilitation Engineering, 1999
- Plantar soft tissue thickness during ground contact in walkingJournal of Biomechanics, 1999
- A rate-controlled indentor for in vivo analysis of residual limb tissuesIEEE Transactions on Rehabilitation Engineering, 1998
- Estimating the effective Young's modulus of soft tissues from indentation tests—nonlinear finite element analysis of effects of friction and large deformationMedical Engineering & Physics, 1997
- In-Shoe Plantar Pressures During Activities of Daily Living: Implications for Therapeutic Footwear DesignFoot & Ankle International, 1996
- A new method for theoretical analysis of static indentation testJournal of Biomechanics, 1996
- An ultrasound indentation system for biomechanical properties assessment of soft tissues in-vivoIEEE Transactions on Biomedical Engineering, 1996
- Intraclass correlations: There's more there than meets the eye.Psychological Bulletin, 1983