Design Considerations and Preliminary Characterizations of a Kilowatt-Plus Microwave-Induced Plasma
- 1 August 1988
- journal article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 42 (6) , 1061-1065
- https://doi.org/10.1366/0003702884430263
Abstract
A high-powered helium microwave-induced plasma was developed, and preliminary characterizations are presented. This kilowatt-plus microwave-induced plasma (KiP-MIP) was contained in a helical flow quartz torch. Power was transferred via waveguides from a 3.0-kW, 2.45-GHz, 120-pulses-per-second microwave generator to a waveguide-to-coaxial converter. A modified Beenakker TM010 resonator cavity was used to focus the energy. Effects of the magnitude of applied power on helium line emission intensities were studied, as well as comparisons of the temporal output of generator output and helium line emission. Introduction of gaseous halogens (Cl2, CBr4) produced intense ion emission which follows the intensity ratios commonly observed in He-MIPs.Keywords
This publication has 16 references indexed in Scilit:
- Moderate-power argon microwave-induced plasma for the detection of metal ions in aqueous samples of complex matrixAnalytical Chemistry, 1987
- Directly coupled chromatography-atomic spectroscopy. Part 1. Directly coupled gas chromatography-atomic spectroscopy. A reviewThe Analyst, 1986
- Tuning and matching the TM1 cavityReview of Scientific Instruments, 1983
- An Internally Tuned TM010 Microwave Resonant Cavity for Moderate Power Microwave-induced PlasmasApplied Spectroscopy, 1983
- Element-selective detection for chromatography by plasma emission spectrometryAnalytica Chimica Acta, 1981
- Microwave-excited atmospheric pressure helium plasma emission detection characteristics in fused silica capillary gas chromatographyAnalytical Chemistry, 1981
- Evaluation of a microwave-induced plasma in helium at atmospheric pressure as an element-selective detector for gas chromatographySpectrochimica Acta Part B: Atomic Spectroscopy, 1977
- A cavity for microwave-induced plasmas operated in helium and argon at atmospheric pressureSpectrochimica Acta Part B: Atomic Spectroscopy, 1976
- Determination of Organophosphorus Insecticide Residues Using the Emission Spectrometric Detector.Analytical Chemistry, 1965
- Sensitive Selective Gas Chromatography Detector Based on Emission Spectrometry of Organic Compounds.Analytical Chemistry, 1965