Enzyme-amplified immunoassays: a new ultrasensitive assay of thyrotropin evaluated.
Open Access
- 1 January 1986
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 32 (1) , 88-92
- https://doi.org/10.1093/clinchem/32.1.88
Abstract
An enzyme-amplified immunoassay for thyrotropin has been evaluated. The lower limit of detection of the assay (mean + 3 SD of the zero standard) was 0.037 milli-int.unit/L and the precision was good, giving a working range of 0.13-24.0 milli-int.units/L. The accuracy of the assay was good, as judged from analytical-recovery experiments, analysis of external quality-assessment samples and comparison with in-house assays. No significant interferences or cross reactivities were identified. Assay of a 25-microL serum sample on the microtiter plate takes 3 h. Values for healthy subjects ranged from 0.4 to 4.0 milli-int.units/L. For 32 thyrotoxic patients thyrotropin concentrations were clearly suppressed and completely distinct from normal values. Patients taking triiodothyronine also showed lower concentrations. The enzyme-amplified immunoassay may thus be a suitable "front line" test for assessing thyroid function.This publication has 6 references indexed in Scilit:
- A sensitive immunoradiometric assay for serum thyroid stimulating hormone: a replacement for the thyrotrophin releasing hormone test?BMJ, 1984
- A rapid, sensitive two-site immunometric assay for TSH using monoclonal antibodies: Investigation of factors affecting optimisationJournal of Immunological Methods, 1984
- Use of a rapid, sensitive immunoradiometric assay for thyrotropin to distinguish normal from hyperthyroid subjects.Clinical Chemistry, 1984
- Thermal gradients in microtitration plates. Effects on enzyme-linked immunoassayJournal of Immunological Methods, 1981
- Outliers in clinical chemistry quality-control schemes.Clinical Chemistry, 1979
- Incorrect least-squares regression coefficients in method-comparison analysis.Clinical Chemistry, 1979