On using a Weissenberg rheogoniometer for oscillatory testing of stiff materials
- 1 November 1974
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 7 (11) , 934-937
- https://doi.org/10.1088/0022-3735/7/11/021
Abstract
When the Weissenberg rheogoniometer is used for oscillatory testing on stiff systems and experimental results interpreted in the conventional way, anomalous results occur. These are due to a lack of rigidity in rheogoniometer parts such as the air-bearing rotor, extension piece and lower platen holder, especially the latter. Future experiments on stiff systems will be facilitated by more rigid instrument members. Existing experimental results can be re-interpreted on the basis of the theory given.Keywords
This publication has 1 reference indexed in Scilit: