Biocompatibility tests of components of an implantable cardiac assist device

Abstract
A permanently implantable in‐series left ventricular assist device, the dynamic aortic patch (DAP), has been tested in chronic animal experiments. The DAP replaces a section of the intrathoracic aortic wall. Hemothorax and hematocele at the implantation site have been complications in recent experiments. Primary postoperative hemorrhage was ruled out, and the biocompatibility of all components was therefore examined. Dacron velour, Teflon felt, conductive polyurethane, segmented polyether polyurethane, and Teflon‐coated polyester fiber sutures were implanted in the pleural cavities of dogs and tested in vitro by culturing canine saphenous vein explants on them. In vivo experiments demonstrated that all components elicited mild to moderate inflammatory reactions, but hematocele occurred only when the components were implanted in the aorta with direct blood contact and exposed to arterial blood pressures. In vitro, cells were cultured on all components with no signs of toxic reactions. These results indicated that the host tolerated all implant components without major inflammatory responses. However, histological data indicated that chronic slow bleeding into or through the Dacron velour in contact with the arterial blood serum could account for hemothorax or hematocele formation. Therefore, a configuration of the assist device using materials impermeable to blood may obviate these difficulties.