Automatic gradient preemphasis adjustment: A 15‐minute journey to improved diffusion‐weighted echo‐planar imaging
Open Access
- 20 December 2001
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 47 (1) , 208-212
- https://doi.org/10.1002/mrm.10022
Abstract
Image distortion caused by gradient eddy currents is a major problem in the use of diffusion tensor imaging (DTI), as using the uncorrected images for calculation of apparent diffusion coefficient (ADC) and diffusion anisotropy will result in areas of artificially increased anisotropy and ADC at the edge of the images, as well as decreased spatial resolution and accuracy in ADC computations overall. This distortion may be substantially reduced by careful adjustment of the gradient preemphasis unit. A completely automatic method of adjusting the preemphasis unit is proposed which finds the optimal settings for all three gradient directions in approximately 15 min by estimating the magnitudes of the eddy currents at various delay times after a test gradient. The pixel shifts in a 64 × 128 echo‐planar diffusion‐weighted image with a diffusion gradient strength of 30 mT/m were reduced to less than 0.2. Magn Reson Med 47:208–212, 2002.Keywords
This publication has 16 references indexed in Scilit:
- Gradient preemphasis calibration in diffusion-weighted echo-planar imagingMagnetic Resonance in Medicine, 2000
- On the use of water phantom images to calibrate and correct eddy current induced artefacts in MR diffusion tensor imagingMagnetic Resonance Imaging, 2000
- Mapping eddy current induced fields for the correction of diffusion-weighted echo planar imagesPublished by Elsevier ,1999
- An Optimized Pulse Sequence for Isotropically Weighted Diffusion ImagingJournal of Magnetic Resonance, 1999
- Localized Eddy Current Compensation Using Quantitative Field MappingJournal of Magnetic Resonance, 1998
- Correction for distortion of echo‐planar images used to calculate the apparent diffusion coefficientMagnetic Resonance in Medicine, 1996
- Quantitative characterization of the eddy current fields in a 40‐cm bore superconducting magnetMagnetic Resonance in Medicine, 1994
- A simple method of measuring gradient induced eddy currents to set compensation networksMagnetic Resonance in Medicine, 1993
- Analysis of the temporal and spatial dependence of the Eddy current fields in a 40‐cm bore magnetMagnetic Resonance in Medicine, 1992
- Improved algorithm for noniterative time-domain model fitting to exponentially damped magnetic resonance signalsJournal of Magnetic Resonance (1969), 1987