Evaluation of a New Ambulatory Spirometer for Measuring Forced Expiratory Volume in One Second and Peak Expiratory Flow Rate
- 1 May 1993
- journal article
- Published by American Thoracic Society in American Review of Respiratory Disease
- Vol. 147 (5) , 1245-1250
- https://doi.org/10.1164/ajrccm/147.5.1245
Abstract
A new pocket-sized ambulatory spirometer (meter) that measures FEV1 and PEFR was evaluated. The meter determines flow rate (V) from the differential pressure (P1-P2) across a sharp-edged orifice (V alpha square root of[P1-P2]). Ten meters were evaluated for accuracy using a computerized syringe to deliver the 24 ATS-recommended waveforms and nine scaled versions of a single waveform (PW#24). Within-meter reproducibility of FEV1 and PEFR was evaluated in two meters by passing five versions of waveform PW#24 through each meter 10 times. Reliability in the ambulatory setting was assessed in six meters on several occasions over a 10-week period using five versions of waveform PW#24. Results show that the 10 meters conform to the ATS accuracy specifications for PEFR with one or less errors and marginally outside these limits for FEV1 with four errors. For the nine versions of PW#24, the 95% confidence intervals indicate that the meter is accurate to within +/- 5.5% or +/- 15 L/min for PEFR and +/- 3.5% or +/- 0.12 L for FEV1. The mean within-meter coefficient of variation was 1.24% for FEV1 and 0.35% for PEFR. There was no significant change in meter accuracy or performance over the 10-wk reliability study. We conclude that the meter is suitable for use as an ambulatory spirometer for measuring FEV1 and PEFR.Keywords
This publication has 15 references indexed in Scilit:
- Evaluation of Accuracy and Reproducibility of Peak Flowmeters at 1,400 mChest, 1992
- Comparison of Peak Expiratory Flow Rate and FEV1 in Assessing Bronchomotor Tone after Challenges with Occupational SensitizersChest, 1991
- Peak expiratory flow at altitude.Thorax, 1990
- Measurement of FEV1 and FVCAnaesthesia, 1988
- Evaluation of a hand-held spirometer, the Respiradyne, for the measurement of forced expiratory volume in the first second (FEV1), forced vital capacity (FVC) and peak expiratory flow rate (PEFR)Respiratory Medicine, 1988
- Evaluation of the turbine pocket spirometer.Thorax, 1987
- An Assessment of Three Portable Peak Flow MetersChest, 1982
- Pocket-sized device for measuring forced expiratory volume in one second and forced vital capacity.BMJ, 1982
- A miniature Wright peak-flow meter.BMJ, 1978
- Maximum Forced Expiratory Flow Rate as a Measure of Ventilatory CapacityBMJ, 1959