Evaluation of the IDI-MRSA assay on the SmartCycler real-time PCR platform for rapid detection of MRSA from screening specimens
- 1 June 2007
- journal article
- research article
- Published by Springer Nature in European Journal of Clinical Microbiology & Infectious Diseases
- Vol. 26 (7) , 459-466
- https://doi.org/10.1007/s10096-007-0303-7
Abstract
Rapid accurate detection is a prerequisite for the successful control of meticillin-resistant Staphylococcus aureus (MRSA). The IDI-MRSA real-time polymerase chain reaction (PCR) assay was designed to provide rapid results from nasal specimens collected in Stuart’s liquid transport medium. This study has evaluated the IDI-MRSA kit for use in a clinical laboratory by investigating the following parameters: (1) limits of detection (LoD), (2) performance with Amies’ gel-based transport medium, (3) ability to detect strains of MRSA in a collection representative of MRSA in Ireland since 1974 (n=113) and (4) performance in a clinical trial with swabs from nose, throat and groin/perineum sites from 202 patients. LoDs (colony-forming units per ml) of the IDI-MRSA kit, direct culture on MRSA-Select chromogenic agar (CA) and salt-enrichment culture (with subculture onto CA) were 103, 103 and 102, respectively. LoDs with Stuart’s and Amies’ transport media were comparable. All except one of the 113 MRSA isolates were detected by the kit but, when six control strains carrying staphylococcal cassette chromosome mec (SCCmec) type IV element subtypes IVa–d and SCCmec types V and VT were tested, the kit failed to detect MRSA carrying SCCmec V. The sensitivity and specificity for detection of MRSA from nose, throat and groin/perineum specimens were comparable with slightly lower sensitivities from throat and groin/perineum specimens compared with nasal swabs (90%, 97%; 89%, 99%; 88%, 99%, respectively). Overall sensitivity, specificity and positive and negative predictive values for specimens from all sites were 88%, 99%, 94% and 97%, respectively. Further developments to improve the sensitivity of this highly worthwhile assay are required.Keywords
This publication has 18 references indexed in Scilit:
- Concurrent Analysis of Nose and Groin Swab Specimens by the IDI-MRSA PCR Assay Is Comparable to Analysis by Individual-Specimen PCR and Routine Culture Assays for Detection of Colonization by Methicillin-Resistant Staphylococcus aureusJournal of Clinical Microbiology, 2006
- Evaluation of MRSA Select, a new chromogenic medium for the detection of nasal carriage of methicillin-resistant Staphylococcus aureusInternational Journal of Antimicrobial Agents, 2006
- Evaluation of the IDI-MRSA Assay for Detection of Methicillin-Resistant Staphylococcus aureus from Nasal and Rectal Specimens Pooled in a Selective BrothJournal of Clinical Microbiology, 2006
- Epidemiological typing of MRSA isolates from blood cultures taken in Irish hospitals participating in the European Antimicrobial Resistance Surveillance System (1999–2003)European Journal of Clinical Microbiology & Infectious Diseases, 2006
- Prospective Comparison of a New Chromogenic Medium, MRSA Select , to CHROMagar MRSA and Mannitol-Salt Medium Supplemented with Oxacillin or Cefoxitin for Detection of Methicillin-Resistant Staphylococcus aureusJournal of Clinical Microbiology, 2006
- Identification of Methicillin‐ResistantStaphylococcus aureusCarriage in Less than 1 Hour during a Hospital Surveillance ProgramClinical Infectious Diseases, 2005
- Detection of Methicillin-Resistant Staphylococcus aureus Directly from Nasal Swab Specimens by a Real-Time PCR AssayJournal of Clinical Microbiology, 2004
- Response of the Clinical Microbiology Laboratory to Emerging (New) and Reemerging Infectious DiseasesJournal of Clinical Microbiology, 2004
- New Real-Time PCR Assay for Rapid Detection of Methicillin- ResistantStaphylococcus aureusDirectly from Specimens Containing a Mixture of StaphylococciJournal of Clinical Microbiology, 2004
- The Detection Of Methicillin Resistance In Staphylococcus AureusJournal of Medical Microbiology, 1969