Work in progress: hybrid temporal-energy subtraction in digital fluoroscopy.
- 1 June 1983
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 147 (3) , 869-874
- https://doi.org/10.1148/radiology.147.3.6342038
Abstract
Initial clinical results using a digital fluoroscopic implementation of the combined time-energy ("hybrid") subtraction technique are described, with emphasis on carotid and renal imaging. Where patient motion artifacts are due to soft-tissue motion alone, hybrid subtraction can remove them. Due to the need for a finite separation time between high- and low-energy pairs, however, the present implementation of the hybrid technique is not completely immune to soft-tissue motion. The intrinsic signal-to-noise ratio of hybrid imaging is less than that of conventional temporal subtraction. However, since the low-energy temporal subtraction images are included in the hybrid data set, the diagnostic quality of the examination is not compromised.Keywords
This publication has 0 references indexed in Scilit: