A prospective comparison between three-dimensional magnetic resonance imaging and ventriculography for target-coordinate determination in frame-based functional stereotactic neurosurgery
- 1 December 1999
- journal article
- research article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 91 (6) , 911-914
- https://doi.org/10.3171/jns.1999.91.6.0911
Abstract
OBJECT: The purpose of this prospective study was to compare stereotactic coordinates obtained with ventriculography with coordinates derived from stereotactic computer-reconstructed three-dimensional magnetic resonance (3D-MR) imaging in functional stereotactic procedures. METHODS: In 15 consecutive patients undergoing functional stereotactic procedures, both preoperative frame-based stereotactic 3D-MR imaging and intraoperative ventriculography were performed. Differences between 3D-MR imaging and ventriculography in X, Y, and Z coordinates of the anterior commissure (AC), posterior commissure (PC), and target area were calculated, as well as the 3D distance between the position of AC, PC, and target within stereotactic space as obtained using both methods. The position of the stereotactic MR imaging fiducial markers measured using 3D-MR imaging compared well with the markers' known position embedded in the software (mean error 0.4 mm, maximal error for an individual slice 1.2 mm). For the individual coordinates, only for Y-PC was a difference found between 3D-MR imaging and ventriculography that significantly exceeded half the size of a pixel, the theoretical limit of precision when using a digitized imaging technique. However, the mean difference was smaller than 1 mm. The mean 3D distance between the 3D-MR imaging- and ventriculography-derived coordinates was 1.09 mm for AC, 1.13 mm for PC, and 1.29 mm for the targets. CONCLUSIONS: With these data it is shown that there is sufficient agreement between ventriculography-derived and 3D-MR imaging-derived stereotactic coordinates to justify the use of 3D-MR imaging target determination in frame-based functional stereotactic neurosurgerKeywords
This publication has 17 references indexed in Scilit:
- Optimizing accuracy in magnetic resonance imaging—guided stereotaxis: a technique with validation based on the anterior commissure—posterior commissure lineJournal of Neurosurgery, 1999
- High-frequency stimulation of the globus pallidus internalis in Parkinson's disease: a study of seven casesJournal of Neurosurgery, 1997
- MRI Distortion and Stereotactic Neurosurgery Using the Cosman-Roberts-Wells and Leksell FramesStereotactic and Functional Neurosurgery, 1996
- Stereotactic Ventrolateral ThalamotomyNeurosurgery, 1995
- Outcome after Stereotactic Thalamotomy for Parkinsonian, Essential, and Other Types of TremorNeurosurgery, 1995
- Magnetic resonance image—directed stereotactic neurosurgery: use of image fusion with computerized tomography to enhance spatial accuracyJournal of Neurosurgery, 1995
- Effects of coregistration of MR to CT images on MR stereotactic accuracyJournal of Neurosurgery, 1995
- Computed Tomography versus Magnetic Resonance Imaging in Stereotactic LocalizationStereotactic and Functional Neurosurgery, 1994
- Long-term suppression of tremor by chronic stimulation of the ventral intermediate thalamic nucleusThe Lancet, 1991
- The long-term results of stereotaxic surgery and L-dopa therapy in patients with Parkinson's diseaseJournal of Neurosurgery, 1980